VIBER X1 ™ is an analog vibration measuring instrument for control and measurement of the mechanical vibration and bearing condition. A flexible instrument that fits in every maintenance engineer's pocket.

Specifications:
Vibration transducer
Accelerometer
Standard 100 mV/g ± 15%
(Adjusted to the instrument)
Input amplitude range
Vibration
Max 20 g RMS
With other sensor up to 200 g
Bearing condition
Max 5 gBC
Dynamic range
60 dB (156,1 Hz)
 
Frequency range
Vibration
2 - 1000 Hz
10 - 1000 Hz
2 - Max Hz 10 - Max Hz
User selectable
Bearing condition
0.5 to 16 kHz
 
Vibration units
g-value, mm/s
User selectable
Amplitude presentation
RMS
Signal processing
Analogous
Accuracy
Vibration
± 3 %
Full scale is 20g for acceleration other units are frequency dependent. All values are related to the normal acceleration
9,80665 m/s2 at sea level and 45 ° latitude
Bearing condition
± 5 %
Over 0,05 gBC
Battery
Rechargeable Lithium 
Capacity for batteries is 2300 mA/h for lithium
Operating time
Min 24 or 12 hours for a full charged battery park
Operating times depends on the intensity on the LED.
Ca 24 hours for Lithium.
External charger
5,0 V Lithium
 
LED display
7 segment
 
Enclosure protection
IP 65
 
Operating temp. range
- 20 to 60 °C
 
Weight
285 gram
Instrument, including battery and transducer
Size
(L x W x H)
125mm x 70mm x 40 mm
OBS. You should avoid using other accelerometers because of instrument calibration. If you change the accelerometer, the instrument must be calibrated to preserve the accuracy.
Features:

•  Total level (mm/s RMS) and bearing condition.
•  Easy to use, start and measure.
•  Dust-and waterproof, for use under tough conditions (IP 65).
•  Interchangeable accelerometer with cable
•  Selectable frequency ranges.
•  Analog instruments with high measurement accuracy and reliability.
•  Provides a good security in assessing when a piece of equipment need to be addressed.

Order Information:
  • Minimum Order: 1ST
Contact Detail
Contact:Chia-Yuan Chuang
Address:36, Ln 196, Chueh-Ming Rd, Kaohsiung City , Taiwan (R.O.C.)
TEL:+886-7-3966681
FAX:+886-7-3966612
Email: info@leadfar.com.tw
service@leadfar.com.tw
ZIP:807